H2 Analyzer – CATARC MP
General description:
The analyzer CATARC MP integrates a sensor which works on the principle of thermal conductibility and a very powerful electronics with microprocessor. It is designed for monitoring and checking binary gas blending like H2/N2, He/N2, SF6/air, etc.
Technical description :
The touch sensitive display allows to user to see very quickly all parameters : gas content, analog output range, temperature control, alarms status, and to get access to the menus.
The instrument integrates some data storing facilities (350 data memory as standard and memory extension with SRAM card as option).
It gives a complete traceability on the events that occur on the analyzer : thresholds alarms, error messages, calibration, configuration, etc. All these data are available on a parallel port (for printer connection) and on a digital interface RS232.
The automatic calibration function is very easy to use and gives the following possibilities : calibration values setting, calibration sequence performed manually, at set time or at regular intervals, calibration of one or two points, safety procedure in case of error on the gas used for calibration.